Wuxi Jinyibo Instrument Technology Co., Ltd.
Main products:Spectrometer
Products
Contact Us
  • Contact Person : Ms. Chen Iry
  • Company Name : Wuxi Jinyibo Instrument Technology Co., Ltd.
  • Tel : 86-510-83217963
  • Fax : 86-510-83217963
  • Address : Jiangsu,Wuxi,No.35 Jingshen Rd, Qianqiao Development Zone
  • Country/Region : China
  • Zip : 214151

CS-8820 Type High-frequency Infrared Carbon & Sulfur Analysis Instrument

CS-8820 Type High-frequency Infrared Carbon & Sulfur Analysis Instrument
Product Detailed
1.C&S ananlysis 2.Double CPU design 3.Time of analysis:20 to 60 seconds 4.Tailor-made new platinum infrared light source

Main technical parameters:

Range of measurement: C: 0.0001%-10.0000% (can be extended to 99.9999%)

                                               S: 0.0001%-3.5000% (can be extended to 99.9999%) 

Analysis error: accord with GB/T20123-2006/ISO15350:2000 standard

Time of analysis: 20 to 60 seconds, could be adjusted. Around 35 seconds usually.

High-frequency furnace: power ≥2.5KVA   Frequency: 18MHz

Electronic balance: weight the sample in variable amount                Accuracy of reading: 0.0001g

Work environment: indoor temperature: 10-30°C

                                            relative humidity:<90%

Power supply: ground connection is good 

                            voltage: AC220V±5%

                            frequency: 50Hz±2%

Oxygen: purity≥99.5%, input pressure: 0.18MPa±5%

Allocation and main features: 

1) Infrared Carbon & Sulfur Analyzer 

Circuit:Adopt double CPU design, electronic circuit integration degree is high,steady and reliable. Adopt multistage hidden isolating circuit and optical transmission, which can settle the high-frequency interference thoroughly. 

Power supply:Industrial integration linear module power supplies, stable output, trouble-free. 

Light source:Tailor-made new platinum infrared light source, fever continued, and high efficiency of the spectral characteristics. 

Analysis pond:Gilded C&S analysis pond, high-precision pyroelectricity infrared detector and high-performance amplifying circuit design, to insure the detection sensitivity and stability. 

2) High -frequency Reaction Burner

High-frequency circuit: high-power High-frequency circuit design, ceramic vacuum capacitance, can burn without samples, lighten the load of High-frequency combustion system, lengthened the service life

Control circuit: automatic detection magnetic valve, go up and down cylinder and high frequency working condition.

Safety system: automatic overtime and overcurrent warning, to protect high-frequency furnace works in normal condition.

Gas circuit structure: gas-entrance circuit system (magnetic valve, pipe joint, go up and down cylinder), automatic leak hunting.

Dust pelletizing system: automatic cleaning device in the furnace end, patent ash-discharge system. Dedusting adopts internal pressure seal, tightness and evolution effect are good, easy to replace, can reduce the influence of dust to analysis results.

Furnace structure: heat in low tension, constant temperature in furnace end, can enhance percent conversion of S; stabilize the analysis result of S.

 

3) Software:

Analysis channel:provide channel management function,C&S channel can add, delete and edit by itself, no quantitative restriction.

Sample management: edit the name, logo of samples, increase and delete sample name.

Show function: C and S hold a curve box respectively; show the real-time data and C, S release curve in the analysis process dynamically.

Data processing function: The analysis results are stored by using the ACCESS database, all the analysis results and curves of C & S can be stored every time.

Inquiry function: the analysis results can be queried according to time, operator, sample name, logo and so on.

Software function: provide curve/data storage, blank deductions, set parameters, channel selection, mathematical statistics, the results correction and curve comparation function etc.

Print function: multiple print mode, the two print modes-laboratory and check station are provided, and it can plan print format by itself.

Self diagnosis function: System diagnosis, detecting sealing of the furnace head and gas chamber by software.

4) Computer and printer allocation:

Microcomputer: brand commercial microcomputer (standard configuration, can be allocated according to customers’ requirements)

Printer: laser printer (A4 paper) (standard configuration, can be allocated according to customers’ requirements)

5) Electronic balance: can be allocated with different brands and different parameter electronic balance according to customers’ requirements, data interface is unitized.

 

 

CS-8820 Type High-frequency Infrared Carbon & Sulfur Analysis Instrument



Copyright Notice @ 2008-2022 ECVERY Limited and/or its subsidiaries and licensors. All rights reserved.