CS-8800C Type High-frequency Infrared Carbon & Sulfur Analysis Instrument

CS-8800C Type High-frequency Infrared Carbon & Sulfur Analysis Instrument
Product Detailed
1.CO translates into CO2 device(only domestic owns) 2.Sensibility (Min. reading): 0.1ppm 3.Double carbon and sulfur ponds

Application fields

Testing the mass fraction of C, S of steel, iron, alloy, mold and cores, nonferrous metal, cement, ore, coke, catalyst, magnetic materials, ceramic, inorganic substance, black lead, refractory matter, battery matter, plant and other materials.

Main technical parameters

Analysis principle

High frequency furnace combustion infrared absorption method

Measuring range

C:0.00001%-10.0000% ( expandable to 99.99999% )

S:0.00001%-6.0000% ( expandable to 99.99999% )

Analysis of error

With GB / T20123-2006/ ISO15350: 2000 standard

Analysis of precision

C: RSD≤0.5 %         S: RSD≤1.0%

Sensitivity ( minimum reading )

0.1ppm

Detecting pool

C: C Pond ( can add low C pond )   S: S Pond ( can add High S pond )

Analysis time

20-100 seconds can be set ( usually 35 seconds)

High frequency furnace

power ≥2.7KVA Oscillation frequency:20MHz

Electronic balance

weight the sample in variable amount Accuracy of reading: 0.0001g

Working environment

indoor temperature:10-30°C,relative humidity: <75% voltage: AC220V±5%,frequency: 50Hz±2%

Key features:

1,humanized circular arc,stainless steel table-board design,easy to operate.

2,domestic initiation double carbon and double sulfur pool access to analysis the ultra low and higher content, and automatic switching.

3,sampling 24 times/s realize the analysis procedure dynamic data integral, improve the sensitivity and accuracy rating of analysis.

4. use7.5KVA high-frequency ceramic power tube, burn the samples adequately and stability is good.

5. Snap-in type automatic sweeper( with patent ) and ultramicropore mental dust filter,reducing the influence of dust to analysis result, easy to dismantle, be in favour of maintain.

6. low tension heating ring( with patent technology) heating in furnace end, which can improve the percent conversion and stabilize the analysis result of S.

7. infrared detection part and high-frequency furnace are connected by optical (optical pioneer),coordinate with multilevel covert style isolation circuit,which can avoid the high-frequency interference thoroughly.

CS-8800C Type High-frequency Infrared Carbon & Sulfur Analysis Instrument



Copyright Notice @ 2008-2022 ECVERY Limited and/or its subsidiaries and licensors. All rights reserved.